scientific research and production company
analytical systems

ASYS ION — Total Hydrocarbon and Oil in Water Analyzer

Technical details Analyzed gases Questionary

Typical applications

The ASYS ION analyzer is designed to detect different type of hydrocarbons in aqueous samples.
- Steam condensate and cooling water control
- Possibility to detect near all type of volatile hydrocarbons
- Different detectors for multiple applications
- Highly effective sample conditioning system for extremely dirty samples

Measuring principle

The analyzer works on completely controlled stripping principle, where water sample flow and temperature and striping gas flow (air or nitrogen) and all main parameters of analyzer are controlled by internal PLC. There are no any dozing valves – it is direct measurement of stripped components on PID, FID, TCD and many other detectors.

The analyzer is a system that allows, by continuous bubbling, to remove hydrocarbons from the water and monitor their concentration using various types of detectors: flame ionization (PID), photo-ionization (FID), semiconductor (PPD), thermo-catalytic (TCD), electrochemical (ECD) and other types of detectors, depending on the type of controlled substances.

The carrier gas is continuously passed through the sample heated to the required temperature, the hydrocarbon molecules from the water fall into the gas and are carried to the detector, which forms the primary measuring signal.

The carrier gas, as a rule, is the instrument air, and for particularly clean environments and for measuring of particularly low concentrations in media such as water vapor condensate, it is advisable to use network nitrogen to exclude the influence of impurities contained in the compressed air.

Based on the calibration data, the controller converts the detector signal into a concentration and generates the analyzer output signal.

Analyzer design

The analyzer structurally consists of two compartments: the electronics compartment and the analytical compartment. In the first case there are controllers, relays and all the necessary electrical equipment for the operation of the device. In the analytical compartment there is a bubbler, a heater and a primary converter of the appropriate type.

The device is implemented on a modern controller base and has extensive self-diagnosis functions. All the main parameters, such as air pressure, temperature in the bubbler, temperature in the device casing and other necessary for the smooth operation of the device are controlled by the device. All the most important parameters are saved on a flash drive.

Analyzer is equipped by 7" LCD.

There are several explosion proof versions, the most convenient are Expx for PID and Expz for other detector types.

The purge control system is located at the panel above the analyzer.

The analyzer design allows easy access for repair and maintenance of all analyzer components.

The following additional functions are available as options:

- Filter back-purge function (auto-cleaning), which is important for polluted environments;
- Automatic sampling for laboratory analysis;
- Automatic analysis of multiple streams.

Our specialists are ready to consider options for upgrading the device in accordance with the measurement tasks and customer requirements.

Technical details

Maximum sample temperature
up to 50 °C (standard) up to +250 °C (with cooler)
Sample flow rate
from 50 to 500 ml / min (standard 400 ml / min)
Operating temperature
5 .. 50 °С
Overall dimensions of the analyzer
from 600x600x300 mm to 1400x1000x600 mm (HxWxD) (depending on the design of the analyzer, detector, type of explosion protection, installed options)
Marking of explosion protection
1Ex px IIC T4 Gb X or 2Ex pz IIC T4 Gc X
Number of measured flows
from 1 to 5 (depending on the length of the lines, samples)
Display
touch LCD 7"
Analog output signals
0/4-20 mA
Relay outputs
of the "dry contact" type by concentration and states
Digital output signals
RS232 / RS485, Ethernet TCP / IP, Modbus, etc
Power supply
220 V, 50 Hz, 300 W (up to 1000 W during warm-up)
Measuring range
FID: minimum 0-5 mg / l, up to 0-1000 mg/l,
PID: minimum 0-1 mg / l, up to 500 mg/l,
TCD: minimum 0-100 mg/l, up to 500 mg/l,
PPD: minimum 0-10 mg/l, up to 1000 mg/l,
ECD: minimum 0-10 mg/L, up to 1000 mg/l
The measurement error
depends on the type of substances being measured
Weight
from 25 to 55 kg (depending on the version)
Degree of casing protection
IP 54 (standard), IP 65 (optional)
Archive memory
SSD removable 8 GB
Air
not less than 300 kPa (2-5 l / min, at start-up up to 100 l / min)

Analytical specification

SubstancesMinimum range
Acetone0 – 10 мг/м3
Benzene0 – 10 мг/м3
Gasoline0 – 10 мг/м3
Butane0 – 10 мг/м3
Butadiene-1,30 – 10 мг/м3
Hydrogen0 – 25 мг/м3
Hexane0 – 10 мг/м3
Heptane0 – 10 мг/м3
Diesel fuel0 – 10 мг/м3
Diethyl ether0 – 10 мг/м3
Isobutylene0 – 10 мг/м3
Kerosene0 – 10 мг/м3
Methane0 – 10 мг/м3
Methyl mercaptan0 - 10 мг/м3
n-Octane0 - 10 мг/м3
Pentane0 - 10 мг/м3
Hydrogen sulfide0 - 10 мг/м3
Styrene0 - 10 мг/м3
Toluene0 - 10 мг/м3
Ethylbenzene0 - 10 мг/м3

* The main substances to be measured are indicated. To get information about your task, you need to fill out a questionnaire.